dc.description.abstractObjectives: This study aimed to compare the mean surgical time, cumulative dissipated energy (CDE), aspiration timeand estimated aspiration fluid parameters in phacoemulsification surgery using an active-fluidic pressure control sys tem (AFPCS) in comparison with gravity-fluidic system (GFS). Methods: The study analysed 202 eyes of 202 patients diagnosed with senile nuclear cataract and admitted to ourclinic. Phacoemulsification surgery was performed by an experienced surgeon using the Centurion Vision System de vice and the Infiniti Vision System device. After each case, the mean surgical time, CDE, aspiration time and estimatedaspiration fluid parameters were recorded. Complicated cases were excluded from the study. Results: In the AFPCS group, the mean age of the patients was 64.15±3.83 years, with the following results: mean surgi cal time, 11.88±5.44 min; CDE, 11.63±9.37; aspiration time, 3.91±1.49 min; estimated aspiration fluid, 71.89±27.56 ml. Inthe GFS group, the mean age of the patients was 64.99±3.80 years, and the following results were obtained: mean sur gical time, 14.35±3.93 min; CDE, 13.60±7.61; aspiration time, 5.56±1.8 min; estimated aspiration fluid, 101.23±15.62 ml. Conclusion: AFPCS was found to be more successful than GFS, with low-energy consumption, shorter surgical timeand better postoperative outcomes.
dc.identifier.citationBulut, A. (2021). Comparison of phacoemulsification parameters using active fluidic pressure control system versus gravity-fluidic syste. Eurasian Journal of Medical Investigation, 5(3), 309-312. https://dx.doi.org/10.14744/ejmi.2021.25231
dc.identifier.doi10.14744/ejmi.2021.25231
dc.identifier.endpage312
dc.identifier.issn2602-3164
dc.identifier.issue3
dc.identifier.startpage309
dc.identifier.trdizinid491615
dc.identifier.urihttps://dx.doi.org/10.14744/ejmi.2021.25231
dc.identifier.urihttps://hdl.handle.net/20.500.12511/10805
dc.identifier.volume5
dc.indekslendigikaynakTR-Dizin
dc.institutionauthorBulut, Asker
dc.language.isoen
dc.publisherKare Publishing
dc.relation.ispartofEurasian Journal of Medical Investigationen_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsAttribution-NonCommercial 4.0 International*
dc.rightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttps://creativecommons.org/licenses/by-nc/4.0/*
dc.subjectActive-Fluidic System
dc.subjectGravity-Fluidic System
dc.subjectPhacoemulsification
dc.subjectCataract
dc.titleComparison of phacoemulsification parameters using active fluidic pressure control system versus gravity-fluidic syste
dc.typeArticle

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
Bulut-Asker-2021.pdf
Boyut:
353.3 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text
Lisans paketi
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: