Hemodynamics of neonatal double lumen cannula malposition

Yükleniyor...
Küçük Resim

Tarih

2020

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

SAGE Publications Ltd.

Erişim Hakkı

info:eu-repo/semantics/embargoedAccess

Özet

Objective: Malposition of dual lumen cannula is a frequent and challenging complication in neonates and plays a significant role in shaping the in vitro device hemodynamics. This study aims to analyze the effect of the dual lumen cannula malposition on right-atrial hemodynamics in neonatal patients using an experimentally validated computational fluid dynamics model. Methods: A computer model was developed for clinically approved dual lumen cannula (13Fr Origen Biomedical, Austin, Texas, USA) oriented inside the atrium of a 3-kg neonate with normal venous return. Atrial hemodynamics and dual lumen cannula malposition were systematically simulated for two rotations (antero-atrial and atrio-septal) and four translations (two intravascular movements along inferior vena cava and two dislodged configurations in the atrium). A multi-domain compartmentalized mesh was prepared to allow the site-specific evaluation of important hemodynamic parameters. Transport of each blood stream, blood damage levels, and recirculation times are quantified and compared to dual lumen cannula in proper position. Results: High recirculation levels (39 ± 4%) in malpositioned cases resulted in poor oxygen saturation where maximum recirculation of up to 42% was observed. Apparently, Origen dual lumen cannula showed poor inferior vena cava blood–capturing efficiency (48 ± 8%) but high superior vena cava blood–capturing efficiency (86 ± 10%). Dual lumen cannula malposition resulted in corresponding changes in residence time (1.7 ± 0.5 seconds through the tricuspid). No significant differences in blood damage were observed among the simulated cases compared to normal orientation. Compared to the correct dual lumen cannula position, both rotational and translational displacements of the dual lumen cannula resulted in significant hemodynamic differences. Conclusion: Rotational or translational movement of dual lumen cannula is the determining factor for atrial hemodynamics, venous capturing efficiency, blood residence time, and oxygenated blood delivery. Results obtained through computational fluid dynamics methodology can provide valuable foresight in assessing the performance of the dual lumen cannula in patient-specific configurations.

Açıklama

The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This study was supported by European Union FP7 CIG-293931- CardioFluidMechanics, TUBITAK 118M369, and TUBITAK 1003 program grants 115E690 and 115E691 (Principle Investigators: Kerem Pekkan and Ece Salihoglu).

Anahtar Kelimeler

Extracorporeal Membrane Oxygenation, Double Lumen Neonatal Cannula, Blood Residence Time, Hemolysis Index, Oxygen Saturation, Computational Fluid Dynamics

Kaynak

Perfusion-uk

WoS Q Değeri

Q4

Scopus Q Değeri

Q1

Cilt

35

Sayı

4

Künye

Jamil, M., Rezaeimoghaddam, M., Çakmak, B., Yıldız, Y., Rasooli, R., Pekkan, K. ... Salihoğlu, E. (2020). Hemodynamics of neonatal double lumen cannula malposition. Perfusion-uk, 35(4), 306-315. http://doi.org/10.1177/0267659119874697