Comparison of two digital intraoral radiography imaging systems as a function of contrast resolution and exposure time

Yükleniyor...
Küçük Resim

Tarih

2020

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Edizioni Minerva Medica

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Background: To compare the image quality of two different digital imaging systems; one photostimulable phosphor plate system (PSP) and a direct digital radiography system with CMOS imaging sensor; via evaluating contrast resolution among four different exposure times. Methods: Endodontically treated incisor teeth embedded in paraffin blocks are aligned next to a 99.5% Al wedge and exposed for 0.8, 0.1,0.125 and 0.16 seconds using both the CMOS and PSP systems. Using ImageJ software, 5 isometric and isogridded ROI from each root filling area and isometric ROI from the Al stepwedge were calculated. Results: Evaluation of the total of 120 images displayed that PSP system produced significantly higher contrast resolution (P<0.05) in regard to pixel values than the CMOS. The CMOS system was non- responsive to increasing dose (P=0.000). Regarding the EqAl values, no significant difference was determined between groups (P>0.05). Conclusions: The contrast resolution was higher using the PSP system. It can be estimated that, filling material will be more obvious under lower doses using PSP.

Açıklama

Anahtar Kelimeler

Radiography, Dental, Technology, Radiologic, Diagnostic Equipment

Kaynak

Minerva Stomatologica

WoS Q Değeri

N/A

Scopus Q Değeri

Q3

Cilt

69

Sayı

3

Künye

Cesur Aydın, K., Demirel, O. ve Özcan, M. (2020). Comparison of two digital intraoral radiography imaging systems as a function of contrast resolution and exposure time. Minerva Stomatologica, 69(3), 148-152. https://dx.doi.org/10.23736/S0026-4970.19.04286-9