Measurement of intraocular pressure with applanation, dynamic contour, and air-puff tonometers: A comparative study in primary open-angle glaucoma and healthy cases

dc.authorid0000-0001-5335-3860
dc.contributor.authorÖzbilen, Kemal Turgay
dc.contributor.authorKocabora, Mehmet Selim
dc.date.accessioned2022-03-02T13:43:48Z
dc.date.available2022-03-02T13:43:48Z
dc.date.issued2020
dc.departmentİstanbul Medipol Üniversitesi, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümü, Göz Hastalıkları Ana Bilim Dalı
dc.description.abstractObjectives: This study aimed to investigate and compare the reliability of Goldmann applanation tonometer (GAT), dynamic contour (DCT), and noncontact (NCT) tonometers in intraocular pressure (IOP) measurement and the affecting parameters in healthy subjects and cases with primary open-angle glaucoma (POAG). Methods: Left eyes of 64 cases (32 males and 32 females) were selected for this prospective, controlled study. Of these cases, 33 had POAG, and 31 were selected as control. IOP measurement was performed using NCT, DCT, and GAT consecutively for each patient, and then central corneal thickness (CCT) was measured. Ocular pulse amplitude (OPA) and all values were recorded. Results: The mean age was 53.36±10 years (31–80 years), and CCT was 561±45 ?. IOP was found as 16.39±3.75 mmHg with GAT, 17.89±3.55 mmHg with DCT, and 15.76±3.49 mmHg with NCT. A significant difference was found between DCT with NCT and GAT. Whereas, a positive correlation was found between CCT with all the three methods used, with DCT as the weakest. While the correlation between all the three methods was excellent, the strongest was found to be between DCT and GAT. Thick corneas affected all the three methods, but DCT was the least affected. While DCT tends to measure higher than both GAT and NCT, this difference decreased as the corneal thickness increased. OPA was found to be 2.56±1.04 mmHg; no statistical difference was found between the groups. A correlation was found between OPA and IOP, and OPA was found to be significantly higher in women. Conclusion: DCT is minimally affected by corneal factors, especially in thin corneas, and shows excellent correlation with GAT. This new-generation digital tonometer can be used safely in glaucoma diagnosis and follow-up
dc.identifier.citationÖzbilen, K. T. ve Kocabora, M. S. (2020). Measurement of intraocular pressure with applanation, dynamic contour, and air-puff tonometers: A comparative study in primary open-angle glaucoma and healthy cases. Beyoglu Eye Journal, 5(3), 178-187. https://doi.org/10.14744/bej.2020.82713
dc.identifier.doi10.14744/bej.2020.82713
dc.identifier.endpage187
dc.identifier.issn2459-1777
dc.identifier.issn2587-0394
dc.identifier.issue3
dc.identifier.pmid35098085
dc.identifier.startpage178
dc.identifier.trdizinidTkRJNE1qZ3dNQT09
dc.identifier.urihttps://doi.org/10.14744/bej.2020.82713
dc.identifier.urihttps://hdl.handle.net/20.500.12511/9073
dc.identifier.volume5
dc.indekslendigikaynakTR-Dizin
dc.indekslendigikaynakPubMed
dc.institutionauthorKocabora, Mehmet Selim
dc.language.isoen
dc.relation.ispartofBeyoglu Eye Journalen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsAttribution-NonCommercial 4.0 International*
dc.rightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttps://creativecommons.org/licenses/by-nc/4.0/*
dc.subjectIntraocular Pressure
dc.subjectOcular
dc.subjectOpen-Angle Glaucoma
dc.subjectOcular Pulse Amplitude
dc.subjectPascal Dynamic Contour tonometer
dc.subjectTonometry
dc.titleMeasurement of intraocular pressure with applanation, dynamic contour, and air-puff tonometers: A comparative study in primary open-angle glaucoma and healthy cases
dc.typeArticle

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
Kocabora-Mehmet Selim-2020.pdf
Boyut:
234.6 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text
Lisans paketi
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: