dc.identifier.citationÖzbilen, K. T. ve Kocabora, M. S. (2020). Measurement of intraocular pressure with applanation, dynamic contour, and air-puff tonometers: A comparative study in primary open-angle glaucoma and healthy cases. Beyoglu Eye Journal, 5(3), 178-187. https://doi.org/10.14744/bej.2020.82713en_US
dc.identifier.issn2459-1777
dc.identifier.issn2587-0394
dc.identifier.urihttps://doi.org/10.14744/bej.2020.82713
dc.identifier.urihttps://hdl.handle.net/20.500.12511/9073
dc.description.abstractObjectives: This study aimed to investigate and compare the reliability of Goldmann applanation tonometer (GAT), dynamic contour (DCT), and noncontact (NCT) tonometers in intraocular pressure (IOP) measurement and the affecting parameters in healthy subjects and cases with primary open-angle glaucoma (POAG). Methods: Left eyes of 64 cases (32 males and 32 females) were selected for this prospective, controlled study. Of these cases, 33 had POAG, and 31 were selected as control. IOP measurement was performed using NCT, DCT, and GAT consecutively for each patient, and then central corneal thickness (CCT) was measured. Ocular pulse amplitude (OPA) and all values were recorded. Results: The mean age was 53.36±10 years (31–80 years), and CCT was 561±45 μ. IOP was found as 16.39±3.75 mmHg with GAT, 17.89±3.55 mmHg with DCT, and 15.76±3.49 mmHg with NCT. A significant difference was found between DCT with NCT and GAT. Whereas, a positive correlation was found between CCT with all the three methods used, with DCT as the weakest. While the correlation between all the three methods was excellent, the strongest was found to be between DCT and GAT. Thick corneas affected all the three methods, but DCT was the least affected. While DCT tends to measure higher than both GAT and NCT, this difference decreased as the corneal thickness increased. OPA was found to be 2.56±1.04 mmHg; no statistical difference was found between the groups. A correlation was found between OPA and IOP, and OPA was found to be significantly higher in women. Conclusion: DCT is minimally affected by corneal factors, especially in thin corneas, and shows excellent correlation with GAT. This new-generation digital tonometer can be used safely in glaucoma diagnosis and follow-upen_US
dc.language.isoengen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.rightsAttribution-NonCommercial 4.0 International*
dc.rights.urihttps://creativecommons.org/licenses/by-nc/4.0/*
dc.subjectIntraocular Pressureen_US
dc.subjectOcularen_US
dc.subjectOpen-Angle Glaucomaen_US
dc.subjectOcular Pulse Amplitudeen_US
dc.subjectPascal Dynamic Contour tonometeren_US
dc.subjectTonometryen_US
dc.titleMeasurement of intraocular pressure with applanation, dynamic contour, and air-puff tonometers: A comparative study in primary open-angle glaucoma and healthy casesen_US
dc.typearticleen_US
dc.relation.ispartofBeyoglu Eye Journalen_US
dc.departmentİstanbul Medipol Üniversitesi, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümü, Göz Hastalıkları Ana Bilim Dalıen_US
dc.authorid0000-0001-5335-3860en_US
dc.identifier.volume5en_US
dc.identifier.issue3en_US
dc.identifier.startpage178en_US
dc.identifier.endpage187en_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.identifier.doi10.14744/bej.2020.82713en_US
dc.institutionauthorKocabora, Mehmet Selim
dc.identifier.trdizinidTkRJNE1qZ3dNQT09en_US
dc.identifier.pmid35098085en_US


Bu öğenin dosyaları:

Thumbnail

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster

info:eu-repo/semantics/openAccess
Aksi belirtilmediği sürece bu öğenin lisansı: info:eu-repo/semantics/openAccess