Measurement of intraocular pressure with applanation, dynamic contour, and air-puff tonometers: A comparative study in primary open-angle glaucoma and healthy cases
Göster/ Aç
Erişim
info:eu-repo/semantics/openAccessAttribution-NonCommercial 4.0 Internationalhttps://creativecommons.org/licenses/by-nc/4.0/Tarih
2020Üst veri
Tüm öğe kaydını gösterKünye
Özbilen, K. T. ve Kocabora, M. S. (2020). Measurement of intraocular pressure with applanation, dynamic contour, and air-puff tonometers: A comparative study in primary open-angle glaucoma and healthy cases. Beyoglu Eye Journal, 5(3), 178-187. https://doi.org/10.14744/bej.2020.82713Özet
Objectives: This study aimed to investigate and compare the reliability of Goldmann applanation tonometer (GAT),
dynamic contour (DCT), and noncontact (NCT) tonometers in intraocular pressure (IOP) measurement and the affecting
parameters in healthy subjects and cases with primary open-angle glaucoma (POAG).
Methods: Left eyes of 64 cases (32 males and 32 females) were selected for this prospective, controlled study. Of these
cases, 33 had POAG, and 31 were selected as control. IOP measurement was performed using NCT, DCT, and GAT
consecutively for each patient, and then central corneal thickness (CCT) was measured. Ocular pulse amplitude (OPA)
and all values were recorded.
Results: The mean age was 53.36±10 years (31–80 years), and CCT was 561±45 μ. IOP was found as 16.39±3.75 mmHg
with GAT, 17.89±3.55 mmHg with DCT, and 15.76±3.49 mmHg with NCT. A significant difference was found between
DCT with NCT and GAT. Whereas, a positive correlation was found between CCT with all the three methods used, with
DCT as the weakest. While the correlation between all the three methods was excellent, the strongest was found to be
between DCT and GAT. Thick corneas affected all the three methods, but DCT was the least affected. While DCT tends
to measure higher than both GAT and NCT, this difference decreased as the corneal thickness increased. OPA was found
to be 2.56±1.04 mmHg; no statistical difference was found between the groups. A correlation was found between OPA
and IOP, and OPA was found to be significantly higher in women.
Conclusion: DCT is minimally affected by corneal factors, especially in thin corneas, and shows excellent correlation with
GAT. This new-generation digital tonometer can be used safely in glaucoma diagnosis and follow-up