Basit öğe kaydını göster

dc.contributor.authorEdsall, Connor
dc.contributor.authorHuynh, Laura
dc.contributor.authorMustafa, Waleed
dc.contributor.authorHall, Timothy L.
dc.contributor.authorYüksel Durmaz, Yasemin
dc.contributor.authorVlaisavljevich, Eli
dc.date.accessioned2024-07-02T08:39:47Z
dc.date.available2024-07-02T08:39:47Z
dc.date.issued2024en_US
dc.identifier.citationEdsall, C., Huynh, L., Mustafa, W., Hall, T. L., Yüksel Durmaz, Y. ve Vlaisavljevich, E. (2024). Nanoparticle-mediated histotripsy using dual-frequency pulsing methods. Ultrasound in Medicine and Biology, 50(8), 1214-1223. http://dx.doi.org/10.1016/j.ultrasmedbio.2024.04.009en_US
dc.identifier.issn0301-5629
dc.identifier.urihttp://dx.doi.org/10.1016/j.ultrasmedbio.2024.04.009
dc.identifier.urihttps://hdl.handle.net/20.500.12511/12687
dc.description.abstractObjective: Nanoparticle-mediated histotripsy (NMH) is a novel ablation method that combines nanoparticles as artificial cavitation nuclei with focused ultrasound pulsing to achieve targeted, non-invasive, and cell-selective tumor ablation. The study described here examined the effect of dual-frequency histotripsy pulsing on the cavitation threshold, bubble cloud characteristics, and ablative efficiency in NMH. High-speed optical imaging was used to analyze bubble cloud characteristics and to measure ablation efficiency for NMH inside agarose tissue phantoms containing perfluorohexane-filled nanocone clusters, which were previously developed to reduce the histotripsy cavitation threshold for NMH. Methods: Dual-frequency histotripsy pulsing was applied at a 1:1 pressure ratio using a modular 500 kHz and 3 MHz dual-frequency array transducer. Optical imaging results revealed predictable, well-defined bubble clouds generated for all tested cases with similar reductions in the cavitation thresholds observed for single-frequency and dual-frequency pulsing. Results: Dual-frequency pulsing was seen to nucleate small, dense clouds in agarose phantoms, intermediate in size of their component frequencies but closer in area to that of the higher component frequency. Red blood cell experiments revealed complete ablations were generated by dual-frequency NMH in all phantoms in <1500 pulses. This result was a significant increase in ablation efficiency compared with the ∼4000 pulses required in prior single-frequency NMH studies. Conclusion: Overall, this study indicates the potential for using dual-frequency histotripsy methods to increase the ablation efficacy of NMH.en_US
dc.language.isoengen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectAblationen_US
dc.subjectCavitationen_US
dc.subjectDual-Frequencyen_US
dc.subjectHistotripsyen_US
dc.subjectNanocone Clustersen_US
dc.subjectNanoparticle-Mediated Histotripsyen_US
dc.subjectNanoparticlesen_US
dc.titleNanoparticle-mediated histotripsy using dual-frequency pulsing methodsen_US
dc.typearticleen_US
dc.relation.ispartofUltrasound in Medicine and Biologyen_US
dc.departmentİstanbul Medipol Üniversitesi, Mühendislik ve Doğa Bilimleri Fakültesi, Biyomedikal Mühendisliği Bölümüen_US
dc.departmentİstanbul Medipol Üniversitesi, Rektörlük, Sağlık Bilim ve Teknolojileri Araştırma Enstitüsüen_US
dc.authorid0000-0003-2776-5807en_US
dc.identifier.volume50en_US
dc.identifier.issue8en_US
dc.identifier.startpage1214en_US
dc.identifier.endpage1223en_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.identifier.doi10.1016/j.ultrasmedbio.2024.04.009en_US
dc.institutionauthorMustafa, Waleed
dc.institutionauthorYüksel Durmaz, Yasemin
dc.identifier.scopus2-s2.0-85193999215en_US
dc.identifier.pmid38797630en_US
dc.identifier.scopusqualityQ1en_US