Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne gold, and reciproc blue nickel-titanium instruments

dc.authorid0000-0001-8656-7101
dc.contributor.authorGündoğar, Mustafa
dc.contributor.authorÖzyürek, Taha
dc.date.accessioned10.07.201910:49:13
dc.date.accessioned2019-07-10T19:58:03Z
dc.date.available10.07.201910:49:13
dc.date.available2019-07-10T19:58:03Z
dc.date.issued2017
dc.departmentİstanbul Medipol Üniversitesi, Diş Hekimliği Fakültesi, Endodonti Ana Bilim Dalı
dc.descriptionWOS: 000405164000027
dc.descriptionPubMed ID: 28527845
dc.description.abstractIntroduction: The purpose of this study was to compare the cyclic fatigue resistances of Reciproc Blue (VDW, Munich, Germany), HyFlex EDM (Coltene/Whaledent, Altstitten, Switzerland), WaveOne Gold (Dentsply Maillefer, Ballaigues, Switzerland), and OneShape (Micro Mega, Besancon, France) single-file NiTi systems. Methods: Thirty Reciproc Blue R25 (25/.08), 30 HyFlex EDM (25/.), 30 WaveOne Gold Primary (25/.07), and 30 OneShape (25/.06) instruments were included in this study. All the instruments were rotated in artificial canals, which were made of stainless steel with an inner diameter of 1.5 mm, a 60 degrees angle of curvature, and radii of curvatures of 5 mm until fracture occurred, and the time to fracture was recorded in seconds using a digital chronometer. The data were analyzed statistically using Kruskal-Wallis and post hoc Dunn tests via SPSS 21.0 software (SPSS Inc, Chicago, IL). The statistical significance level was set at 5%. Results: The HyFlex EDM file (3456.33 +/- 633.37) file had the statistically highest fatigue resistance, and the OneShape file (1221.63 +/- 812.4) had the least fatigue resistance (P < .05). The mean number of cycles to fracture of the Reciproc Blue file (2875.89 +/- 105.35) file was statistically higher than the WaveOne Gold file (1737.00 +/- 376.32) (P < .05). There was no statistically significant difference (P > .05) in the mean length of the fractured fragments of the files (P > .05). Conclusion: Within the limitations of the present in vitro study, it was found that cyclic fatigue resistance of HyFlex EDM files was higher than the cyclic fatigue resistances of OneShape, Reciproc Blue, and WaveOne Gold files.
dc.identifier.citationGündoğar, M. ve Özyürek, T. (2017). Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne gold, and reciproc blue nickel-titanium instruments. Journal of Endodontics, 43(7), 1192-1196. https://dx.doi.org/10.1016/j.joen.2017.03.009
dc.identifier.doi10.1016/j.joen.2017.03.009
dc.identifier.endpage1196
dc.identifier.issn0099-2399
dc.identifier.issn1878-3554
dc.identifier.issue7
dc.identifier.scopusqualityQ1
dc.identifier.startpage1192
dc.identifier.urihttps://dx.doi.org/10.1016/j.joen.2017.03.009
dc.identifier.urihttps://hdl.handle.net/20.500.12511/3098
dc.identifier.volume43
dc.identifier.wosqualityQ1
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherElsevier Science Inc
dc.relation.ispartofJournal of Endodonticsen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/embargoedAccess
dc.subjectArtificial Canals
dc.subjectCyclic Fatigue Resistance
dc.subjectEndodontics
dc.subjectNickel-Titanium
dc.subjectSingle-File Systems
dc.titleCyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne gold, and reciproc blue nickel-titanium instruments
dc.typeArticle

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
gundogar-mustafa(2017).pdf
Boyut:
872.85 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text