Comparision of optical low coherence reflectometry versus ultrasonic biometry in high hypermetropia

dc.authorid0000-0003-0668-3749
dc.authorid0000-0001-9829-7268
dc.contributor.authorAydın, Rukiye
dc.contributor.authorKaraman Erdur, Sevil
dc.contributor.authorŞerefoğlu Çabuk, Kübra
dc.contributor.authorKarahan, Eyüp
dc.contributor.authorKaynak, Süleyman
dc.date.accessioned10.07.201910:49:13
dc.date.accessioned2019-07-10T19:56:00Z
dc.date.available10.07.201910:49:13
dc.date.available2019-07-10T19:56:00Z
dc.date.issued2018
dc.departmentİstanbul Medipol Üniversitesi, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümü, Göz Hastalıkları Ana Bilim Dalı
dc.descriptionWOS: 000457837000021
dc.descriptionPubMed ID: 27898517
dc.description.abstractPurpose: To compare anterior chamber depth (ACD), axial length (AL), and lens thickness (LT) measurements obtained by the Lenstar LS 900 (Haag-Streit AG) optical low-coherence reflectometry with those obtained by the A-scan contact ultrasound among patients with moderate and high hypermetropia. Methods: Fifty-two eyes of 52 patients with moderate and high hypermetropia (spherical equivalent of +4 D or more) were examined in this study measurements of ACD, AL, and LT obtained by Lenstar were compared with those obtained by applanation A-scan ultrasound. All measurements were obtained by two independent examiners. The interdevice agreements were evaluated with Bland-Altman analyses. Results: The mean age of the patients was 54.78 +/- 12.77 years (range 18-74 years). The mean spherical equivalent refractive power was +5.16 +/- 1.12 D (+4.0 to +8.75). The mean values of ACD, AL, and LT with A-scan were 3.05 +/- 0.34, 21.55 +/- 0.75, and 4.33 +/- 0.49 mm, respectively, whereas these values were 2.99 +/- 0.45, 21.58 +/- 0.78, and 4.20 +/- 0.44 mm, respectively, with Lenstar. There was statistically significant difference of LT between the two methods (P=0.02). The mean differences (lower/upper limit of agreement) of the ACD, AL, and LT values for A-scan ultrasound and Lenstar were -0.06 (-0.594/0.474), 0.04 (-0.380/0.459), and -0.12 (-0.739/0.502), respectively. Conclusion: Among patients with moderate and high hypermetropia, the biometric measurements of ACD, AL, and LT by ultrasound and optical biometry were determined to be correlated and there was a high degree of agreement between contact A-scan ultrasonic biometry and Lenstar.
dc.identifier.citationAydın, R., Karaman Erdur, S., Şerefoğlu Çabuk, K., Karahan, E. ve Kaynak, S. (2018). Comparision of optical low coherence reflectometry versus ultrasonic biometry in high hypermetropia. Eye & Contact Lens-Science and Clinical Practice, 44(Supplement: 1), S115-S117. https://dx.doi.org/10.1097/ICL.0000000000000350
dc.identifier.doi10.1097/ICL.0000000000000350
dc.identifier.endpageS117
dc.identifier.issn1542-2321
dc.identifier.issn1542-233X
dc.identifier.issueSupplement: 1
dc.identifier.scopusqualityQ2
dc.identifier.startpageS115
dc.identifier.urihttps://dx.doi.org/10.1097/ICL.0000000000000350
dc.identifier.urihttps://hdl.handle.net/20.500.12511/2516
dc.identifier.volume44
dc.identifier.wosqualityQ3
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherLippincott Williams & Wilkins
dc.relation.ispartofEye & Contact Lens-Science And Clinical Practiceen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/embargoedAccess
dc.subjectHypermetropia
dc.subjectA-scan Ultrasound
dc.subjectOptical Low Coherence Reflectometry
dc.subjectBiometry
dc.titleComparision of optical low coherence reflectometry versus ultrasonic biometry in high hypermetropia
dc.typeArticle

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
Erdur Karaman, S..pdf
Boyut:
123.8 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text