Angio-seal used as a bailout for incomplete hemostasis after dual perclose ProGlide deployment in transcatheter aortic valve implantation

Yükleniyor...
Küçük Resim

Tarih

2022

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Texas Heart Institute

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Background: The failure rate of vascular closure devices remains a significant cause of major vascular complications in contemporary transcatheter aortic valve implantation practice. Methods: This research aimed to evaluate use of the Angio-Seal device in a bailout context in the setting of incomplete hemostasis following use of dual Perclose ProGlide devices in patients undergoing transfemoral transcatheter aortic valve implantation. A total of 185 patients undergoing transfemoral transcatheter aortic valve implantation with either dual Per-close ProGlide (n = 139) or a combination of dual Perclose ProGlide and Angio-Seal (n = 46) were retrospectively analyzed. The baseline, procedural characteristics, and all outcomes (defined according to Valve Academic Research Consortium-2 criteria) were compared. Results: No significant differences were seen between the dual Perclose ProGlide vs dual Perclose ProGlide+Angio-Seal groups with regard to the in-hospital Valve Academic Research Consortium-2 primary end points of major vascular complications (n = 13 [9.4%] vs n = 2 [4.3%]; P =.36), minor vascular complications (n = 13 [9.4%] vs n = 8 [14.7%]; P =.14), major bleeding (n = 16 [11.5%] vs n = 2 [4.3%]; P =.25), and minor bleeding (n = 9 [6.5%] vs n = 5 [10.9%]; P =.34), with higher rates of hematoma in the dual Perclose ProGlide+Angio-Seal group (n = 4 [2.9%] vs n = 5 [10.9%]; P =.044). Conclusion: Finding from the current study suggest that adjunctive Angio-Seal deployment may be feasible and safe, especially in patients with incomplete hemostasis following dual Perclose ProGlide use, and can be an optimal “bailout” procedure. (Tex Heart Inst J. 2022;49(6):e217684).

Açıklama

Anahtar Kelimeler

Peripheral Artery Diseases, Hemostasis, Vascular Access Device

Kaynak

Texas Heart Institute Journal

WoS Q Değeri

Q4

Scopus Q Değeri

Q3

Cilt

49

Sayı

6

Künye

Çakal, B., Çakal, S., Karaca, O., Kızılırmak Yılmaz, F., Güneş, H. M., Yıldırım, A. ... Boztosun, B. (2022). Angio-seal used as a bailout for incomplete hemostasis after dual perclose ProGlide deployment in transcatheter aortic valve implantation. Texas Heart Institute Journal, 49(6). https://doi.org/10.14503/THIJ-21-7684