The analysis of different percentage depth dose curves used for commissioning in eclipse

dc.authorid0000-0002-3936-080X
dc.authorwosidQ-6082-2019
dc.contributor.authorCebe, Mehmet
dc.contributor.authorAcar, Hilal
dc.contributor.authorKüçük, Nadir
dc.contributor.authorKüçükmorkoç, Esra
dc.contributor.authorDoyuran, Mine
dc.contributor.authorÇağlar, Mustafa
dc.contributor.authorAltınok, Ayşe
dc.contributor.authorYilidir, G.
dc.contributor.authorMabhouti, H.
dc.contributor.authorÇağlar, Hale
dc.date.accessioned2022-04-12T08:31:33Z
dc.date.available2022-04-12T08:31:33Z
dc.date.issued2015
dc.departmentİstanbul Medipol Üniversitesi, Tıp Fakültesi, Dahili Tıp Bilimleri Bölümü, Radyasyon Onkolojisi Ana Bilim Dalı
dc.description.abstractPurpose: Percentage Depth Doses (PDDs) used for balance fitting of the calculation model in Eclipse. The effect of PDD curves measured with two different detectors for the commissioning of Truebeam STx machine to the improvement of the quality of the algorithm configuration is analysed. Methods: The measurement of open field PDD for 2×2cm2 to 40×40cm2 field sizes were done with two different detector types. The scan was made using 0.125cc semiflex thimble ionization chamber (PTW TM31010) and PTW 60019 microDiamond with 0.004mm3 sensitive volume. Two different machines were created in Eclipse with the same beam data except for PDD measured with two different detector types. A single 6 FFF beam with zero gantry and collimator angles for 2×2, 3×3, 5×5cm2field sizes were planned on the CT scan of RW3 solid water. The dose calculation was done with 1mm grid size using AcurosXB v13 algorithm using two different machine created. PDD measurements were made in a RW3 solid water phantom using EBT3 Gafchromic film extending from surface to 30cm. The gamma index analysis of measured and calculated PDD curves were compared. Result: For the plan made using machine configured with semiflex measured PDD curves, the measured and the calculated dose distributions for 2×2cm2 field size gamma analysis passing rate was 97.6% within 2% dose difference and 2mm distance-to-agreement. For the plan made using machine configured with microDiamond measured PDD curves had 98.1% passing rate for ±2% in field but the passing rates. Passing rates were increasing with the field sizes. For 5×5cm2 field size they are 98.3% and 99.1% respectively. Conclusion: The agrement with treatment planning system and measurement was better when fine resolution detector was used especially for small field sizes.
dc.description.sponsorshipAmerican Association of Physicists in Medicineen_US
dc.identifier.citationCebe, M., Acar, H., Küçük, N., Küçükmorkoç, E., Doyuran, M., Çağlar, M. ... Çağlar, H. (2015). The analysis of different percentage depth dose curves used for commissioning in eclipse. 57th Annual Meeting and Exhibition of the American-Association-of-Physicists-in-Medicine (AAPM) içinde (3504-3504 ss.). https://doi.org/10.1118/1.4925090
dc.identifier.doi10.1118/1.4925090
dc.identifier.endpage3504
dc.identifier.issn0094-2405
dc.identifier.issue6
dc.identifier.startpage3504
dc.identifier.urihttps://doi.org/10.1118/1.4925090
dc.identifier.urihttps://hdl.handle.net/20.500.12511/9301
dc.identifier.volume42
dc.identifier.wosWOS:000356998302432en_US
dc.identifier.wosqualityQ1
dc.indekslendigikaynakWeb of Science
dc.institutionauthorCebe, Mehmet
dc.institutionauthorAcar, Hilal
dc.institutionauthorKüçük, Nadir
dc.institutionauthorKüçükmorkoç, Esra
dc.institutionauthorDoyuran, Mine
dc.institutionauthorÇağlar, Mustafa
dc.institutionauthorAltınok, Ayşe
dc.institutionauthorYilidir, G.
dc.institutionauthorMabhouti, H.
dc.institutionauthorÇağlar, Hale
dc.language.isoen
dc.publisherWiley
dc.relation.ispartof57th Annual Meeting and Exhibition of the American-Association-of-Physicists-in-Medicine (AAPM)en_US
dc.relation.publicationcategoryKonferans Öğesi - Uluslararası - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectEclipse
dc.subjectPercentage Depth Doses (PDDs)
dc.subjectTruebeam STx Machine
dc.titleThe analysis of different percentage depth dose curves used for commissioning in eclipse
dc.typeConference Object

Dosyalar

Lisans paketi
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: