A novel model of locking plate and locking spongious screw: a biomechanical in vitro comparison study with classical locking plate

Yükleniyor...
Küçük Resim

Tarih

2024

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Erişim Hakkı

Attribution 4.0 International
info:eu-repo/semantics/openAccess

Özet

Background: Locking plates are commonly used for the fixation of comminuted, periprosthetic and osteoporotic bone fractures. These plates are secured to the bone with screws, creating a stable connection with fixed angle between the plate and the screws. In this biomechanical in vitro study, our aim is to evaluate and compare the novel locking plate-locking spongious screw model with FDA approved classical locking plate. Methods: Sawbone PCF-15 osteoporotic bone model was utilized to simulate osteoporotic bone conditions. Two screws were used to attach both the classical locking plate and the novel locking plate-locking spongious screw model to these bone models. The attachment strength of the screws to the bone blocks was measured by pull-out tests. Results: Novel locking plate-locking spongious screw model exhibited an 84.38% stronger attachment to the osteoporotic bone model compared to the current locking plate model. Conclusions: In conclusion, one of the important problems in the locking plates which is the high Pull-out risk of the locking spongious screws can been resolved with our proposed new model and has a chance of having a better purchase especially in osteoporotic bones.

Açıklama

Anahtar Kelimeler

Locking Plates, Plate Fixation, Pull-Out Failure, Screw Failure

Kaynak

Journal of Orthopaedic Surgery and Research

WoS Q Değeri

Q2

Scopus Q Değeri

Q1

Cilt

19

Sayı

1

Künye

Parmaksızoğlu, F., Kılıç, S. ve Çetin, O. (2024). A novel model of locking plate and locking spongious screw: a biomechanical in vitro comparison study with classical locking plate. Journal of Orthopaedic Surgery and Research, 19(1). http://dx.doi.org/10.1186/s13018-024-04700-4