Effects of ultrasound frequency on nanodroplet-mediated histotripsy

dc.authorid0000-0003-2776-5807
dc.contributor.authorVlaisavljevich, Eli
dc.contributor.authorAydın, Ömer
dc.contributor.authorYüksel Durmaz, Yasemin
dc.contributor.authorLin, Kuangwei
dc.contributor.authorFowlkes, Brian
dc.contributor.authorElsayed, Mohamed
dc.contributor.authorXu, Zhen
dc.date.accessioned10.07.201910:49:13
dc.date.accessioned2019-07-10T20:04:19Z
dc.date.available10.07.201910:49:13
dc.date.available2019-07-10T20:04:19Z
dc.date.issued2015
dc.departmentİstanbul Medipol Üniversitesi, Mühendislik ve Doğa Bilimleri Fakültesi, Biyomedikal Mühendisliği Bölümü
dc.descriptionWOS: 000361899900007
dc.descriptionPubMed ID: 25959056
dc.description.abstractNanodroplet-mediated histotripsy (NMH) is a targeted ultrasound ablation technique combining histotripsy with nanodroplets that can be selectively delivered to tumor cells for targeted tumor ablation. In a previous study, it was reported that by use of extremely short, high-pressure pulses, histotripsy cavitation bubbles were generated in regions containing nanodroplets at significantly lower pressure (similar to 10.8 MPa) than without nanodroplets (similar to 28 MPa) at 500 kHz. Furthermore, it was hypothesized that lower frequency would improve the effectiveness of NMH by increasing the size of the focal region, increasing bubble expansion, and decreasing the cavitation threshold. In this study, we investigated the effects of ultrasound frequency (345 kHz, 500 kHz, 1.5 MHz, and 3 MHz) on NMH. First, the NMH cavitation threshold was measured in tissue phantoms with and without nanodroplets, with results indicating that the NMH threshold was significantly below the histotripsy intrinsic threshold at all frequencies. Results also indicated that the NMH threshold decreased at lower frequency, ranging from 7.4 MPa at 345 kHz to 13.2 MPa at 3 MHz. In the second part of this study, the effects of frequency on NMH bubble expansion were investigated, with results indicating larger expansion at lower frequency, even at a lower pressure. In the final part of this study, the ability of perfluoropentane-encapsulated nanodroplets to act as sustainable cavitation nuclei over multiple pulses was investigated, with results indicating that the nanodroplets are destroyed by the cavitation process and only function as cavitation nuclei for the first few pulses, with this effect being most pronounced at higher frequencies. Overall, the results of this study support our hypothesis that using a lower frequency will improve the effectiveness of NMH by increasing the size of the focal region, increasing bubble expansion and decreasing the cavitation threshold.
dc.description.sponsorshipNational Science Foundation Graduate Research Fellowship; Turkish Republic the Ministry of National Education Fellowship Program [1416]; U.S. Department of Defense [W81 XWH-11-PCRP-ID]en_US
dc.identifier.citationVlaisavljevich, E., Aydın, Ö., Yüksel Durmaz, Y., Lin, K., Fowlkes, B., Elsayed, M. ve Xu, Z. (2015). Effects of ultrasound frequency on nanodroplet-mediated histotripsy. Ultrasound in Medicine and Biology, 41(8), 2135-2147. https://dx.doi.org/10.1016/j.ultrasmedbio.2015.04.007
dc.identifier.doi10.1016/j.ultrasmedbio.2015.04.007
dc.identifier.endpage2147
dc.identifier.issn0301-5629
dc.identifier.issn1879-291X
dc.identifier.issue8
dc.identifier.scopusqualityQ1
dc.identifier.startpage2135
dc.identifier.urihttps://dx.doi.org/10.1016/j.ultrasmedbio.2015.04.007
dc.identifier.urihttps://hdl.handle.net/20.500.12511/4031
dc.identifier.volume41
dc.identifier.wosqualityQ1
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherElsevier Science
dc.relation.ispartofUltrasound in Medicine and Biologyen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectNanodroplet-Mediated Histotripsy
dc.subjectFrequency
dc.subjectCavitation
dc.titleEffects of ultrasound frequency on nanodroplet-mediated histotripsy
dc.typeArticle

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
durmaz, yasemin-2015.pdf
Boyut:
1.65 MB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text