Basit öğe kaydını göster

dc.contributor.authorAnıl Aslan, Cemile
dc.contributor.authorEliaçık, Mustafa
dc.contributor.authorKaraman Erdur, Sevil
dc.contributor.authorŞentürk, Fevzi
dc.contributor.authorGülkılık, Gökhan
dc.date.accessioned2020-08-12T12:38:42Z
dc.date.available2020-08-12T12:38:42Z
dc.date.issued2016en_US
dc.identifier.citationAnıl Aslan, C., Eliaçık, M., Karaman Erdur, S., Şentürk, F. ve Gülkılık, G. (2016). 1-5 yaş arası çocuklarda aydınlık ve karanlık ortamlarda welch allyn sure sight otorefraktometre ölçümleri. Türkiye Klinikleri Oftalmoloji Dergisi, 25(3), 158-163. https://dx.doi.org/10.5336/ophthal.2015-48416en_US
dc.identifier.issn1300-0365
dc.identifier.issn2146-9008
dc.identifier.urihttps://dx.doi.org/10.5336/ophthal.2015-48416
dc.identifier.urihttps://hdl.handle.net/20.500.12511/5702
dc.description.abstractAmaç: Aydınlık ve karanlık ortamlarda pupilde meydana gelen fizyolojik değişimlerin Welch Allyn SureSight (WASS; Welch Allyn, Skaneateles Falls, New York) el otorefraktometresi ile elde edilen ölçümler üzerine olan etkilerini incelemek. Gereç ve Yöntemler: 1-5 yaş grubunda herhangi bir oküler patolojisi bulunmayan 57 çocuk çalışma kapsamına alındı. WASS ile sırasıyla sikloplejisiz aydınlık ve karanlık ortamda ve sonrasında sikloplejili olarak refraktif değerler ölçüldü. Otorefraktometre ölçüleri sonrasında deneyimli pediatrik oftalmolog tarafından tüm hastaların sikloplejili retinoskopik muayeneleri gerçekleştirildi. Analiz için hasta yaşı ve hastaların sağ gözlerinin refraktif (sferik, silindirik, aks) ölçümleri kayıt altına alındı.Bulgular: Çalışmaya dahil edilen hastaların WASS ile %76’sından ve retinoskopi ile %96’sından sağlıklı ölçüm alınabildi. Normal oda aydınlığında ve karanlık ortamda el otorefraktometresi ile alınan sferik eşdeğer ortalaması sırasıyla 0,97±0,60 D (-0,63-2,38) ve 1,85±0,59 D (0,38-3,50) idi. Sikloplejili ortalama sferik eşdeğeri 1,95±0,95 (-0,35-3,50) D iken sikloplejili retinoskopik muayenede elde edilen ortalama sferik eşdeğeri 2,23±0,78 (1,00-4,00) D. Sikloplejisiz aydınlık ortamda tespit edilen sferik ölçümleri ile diğer tüm ölçümler arasında istatistiksel olarak anlamlı fark bulunmaktaydı (p=0,001). Silindirik değerler karşılaştırıldığında sikloplejisiz aydınlık ortam ölçümleri ile diğer ölçümler arasında anlamlı farka rastlanılmadı (p=0,992, 0,557 ve 0,603). Sikloplejili WASS ve skiaskopi ölçümlerine ait Blant Altman analizi ile elde edilen sferik ekivalan fark dağılım ortalaması -0,27 D bulundu. Sonuç: Aydınlık ortamda sikloplejisiz WASS otorefraktometre ile elde edilen sferik değerlerin güvenirliğinin düşük olduğu tespit edildi. Dolayısı ile bu çalışmada otomatik cihazların ancak tarama testi amaçlı kullanılabileceği, altın standardın sikloplejili skiaskopi/retinoskopi olduğu vurgulanmıştır.en_US
dc.description.abstractObjective: To investigate the impact of physiologic pupil changes on the accuracy of the Welch Allyn SureSight (WASS; Welch Allyn, Skaneateles Falls, New York) hand-held autorefractor in photopic and scotopic conditions. Materyal and Method: Data are reported for 57 children, aged 1- to 5-years-old, with no ocular abnormalities. Each of the children was tested in the following order: WASS non-cycloplegic refraction in photopic and scotopic conditions, WASS cycloplegic refraction and cycloplegic retinoscopic refraction by an experienced pediatric ophthalmologist. Analysis variables included age (months) and refraction (sphere, cylinder, axis) of right eye. Results: Data were successfully obtained on 76% of children using WASS and 96% with retinoscopy. The mean WASS non-cyloplegic spherical equivalent (SE) in photopic and scotopic conditions were 0.97±0.60 (range:-0.63 to 2.38) D and 1.85±0.59 D (range: 0.38 to 3.50), respectively. The mean WASS cyloplegic SE was 1.95±0.95 D (range:-0.35 to 3.50). The mean SE obtained by cyloplegic retinoscopy was 2.23±0.78 (range: 1.00 to 4.00) diopters (D). WASS non-cyloplegic refraction in photopic conditions were significantly different from other measurements (p <0.001 for all). There was no difference in cylindrical values among WASS non-cyloplegic refraction in photopic and scotopic conditions, WASS cyloplegic refraction and retinoscopy (p=0.992, 0.557, and 0.603, respectively). Conclusion: The reliability of the noncyclopleged autorefractor spheric measurements with the WASS was low in photopic condition. Therefore, in this study it was emphasized that autorefractometers were useful as screening tools and the gold standard technique was manual retinoscopy.en_US
dc.language.isoturen_US
dc.publisherTürkiye Kliniklerien_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectRetinoskopien_US
dc.subjectRefraksiyonen_US
dc.subjectOküleren_US
dc.subjectRetinoscopyen_US
dc.subjectRefractionen_US
dc.subjectOcularen_US
dc.title1-5 yaş arası çocuklarda aydınlık ve karanlık ortamlarda welch allyn sure sight otorefraktometre ölçümlerien_US
dc.title.alternativeWelch allyn suresight autorefractometer measurements in photopic and scotopic conditions in children aged 1-5 yearsen_US
dc.typearticleen_US
dc.relation.ispartofTürkiye Klinikleri Oftalmoloji Dergisien_US
dc.departmentİstanbul Medipol Üniversitesi, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümü, Göz Hastalıkları Ana Bilim Dalıen_US
dc.authorid0000-0002-2563-2149en_US
dc.authorid0000-0001-9829-7268en_US
dc.authorid0000-0002-8851-6559en_US
dc.identifier.volume25en_US
dc.identifier.issue3en_US
dc.identifier.startpage158en_US
dc.identifier.endpage163en_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.identifier.doi10.5336/ophthal.2015-48416en_US


Bu öğenin dosyaları:

Thumbnail

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster