Basit öğe kaydını göster

dc.contributor.authorGündoğar, Mustafa
dc.date.accessioned2021-02-08T07:03:01Z
dc.date.available2021-02-08T07:03:01Z
dc.date.issued2020en_US
dc.identifier.citationGündoğar, M. (2020). HyFlex EDM, RC Gold ve G-Star nikel titanyum eğelerinin döngüsel yorgunluğa karşı dirençlerinin karşılaştırılması. Türkiye Klinikleri Diş Hekimliği Bilimleri Dergisi, 26(1), 21-25. https://doi.org/10.5336/dentalsci.2018-64545en_US
dc.identifier.issn1300-7734
dc.identifier.issn2146-8966
dc.identifier.urihttps://doi.org/10.5336/dentalsci.2018-64545
dc.identifier.urihttps://hdl.handle.net/20.500.12511/6536
dc.description.abstractAmaç: Statik model altında HyFlex EDM, RC Gold ve G-Star nikel titanyum eğelerinin döngüsel yorgunluğa karşı dirençlerinin karşılaştırılmasıdır. Gereç ve Yöntemler: Yirmi G-Star (25/,06), 20 RC Gold File (25/,08) ve 20 HyFlex EDM (25/,08) çalışmaya dâhil edildi. Eğeler üretici firma talimatlarına uygun olarak 5 mm kurvatür yarı çapına, 60° kanal kurvatür açısına ve 1,5 mm kanal iç çapına sahip paslanmaz çelikten yapılmış yapay kanallarda kırılıncaya kadar kullanıldı. Eğeler kırılıncaya kadar geçen süre dijital kronometre ile kaydedildi ve eğelerin kırılıncaya kadar yaptığı tur sayısı hesaplandı. Elde edilen veriler Kruskal-Wallis testi ile istatistiksel olarak değerlendirildi. Bulgular: Eğelerin kırılıncaya kadar yaptığı ortalama tur sayılarının yüksekten düşüğe sıralaması sırayla HyFlex EDM, RC Gold ve G-Star olarak bulundu. Bütün gruplar arasında istatistiksel olarak fark mevcuttu (p<0,05). Sonuç: Çalışmamız sınırları dâhilinde HEDM eğesinin döngüsel yorgunluğa karşı direnci, RC Gold File ve G-Star eğesinden daha yüksek bulunmuştur.en_US
dc.description.abstractObjective: To compare the cyclic fatigue resistance of HyFlex EDM, RC Gold and G-Star nickel titanium files under static model. Material and Methods: Twenty G-Star (25/.06), 20 RC Gold File (25/.08) and 20 HyFlex EDM (25/,08) nickel titanium files were included in the present study. According to the manufacturer instruction files were rotated in an artificial stainless steel with 5 mm radius, 60⑤ angle of curvature and 1.5 mm inner diameter. The time to failure of files was recorded with a digital chronometers and the number of cycles to failure of files were calculated. Kruskal-Wallis test was performed for statistically analyze. Results: The mean number of cycles to failure of files highest to lowest was HyFlex EDM, RC Gold and G-Star respectively. A statistically significant difference was noted between all the groups (p<0.05). Conclusion: Within the limitation of the present study the cyclic fatigue resistance of HyFlex EDM nickel titanium files was greater than the RC Gold and G-Star nickel titanium files.en_US
dc.language.isoturen_US
dc.publisherTürkiye Kliniklerien_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectEndodontien_US
dc.subjectTitanyum Nikeliden_US
dc.subjectEndodonticsen_US
dc.subjectTitanium Nickelideen_US
dc.titleHyFlex EDM, RC Gold ve G-Star nikel titanyum eğelerinin döngüsel yorgunluğa karşı dirençlerinin karşılaştırılmasıen_US
dc.title.alternativeComparison of cyclic fatigue resistance of HyFlex EDM, RC Gold and G-Star nickel titanium filesen_US
dc.typearticleen_US
dc.relation.ispartofTürkiye Klinikleri Diş Hekimliği Bilimleri Dergisien_US
dc.departmentİstanbul Medipol Üniversitesi, Diş Hekimliği Fakültesi, Endodonti Ana Bilim Dalıen_US
dc.authorid0000-0001-8656-7101en_US
dc.identifier.volume26en_US
dc.identifier.issue1en_US
dc.identifier.startpage21en_US
dc.identifier.endpage25en_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.identifier.doi10.5336/dentalsci.2018-64545en_US


Bu öğenin dosyaları:

Thumbnail

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster